An optimized set-up for helmet noninvasive ventilation improves pressure support delivery and patient–ventilator interaction

Francesco Mojoli| Giorgio A. Iotti| Ilaria Currò| Marco Pozzi| Gabriele Via| Aaron Venti| Antonio Braschi
Original
Volume 39, Issue 1 / January , 2013

Pages 38 - 44

Abstract

Objective

To test the effects on mechanical performance of helmet noninvasive ventilation (NIV) of an optimized set-up concerning the ventilator settings, the ventilator circuit and the helmet itself.

Subjects and methods

In a bench study, helmet NIV was applied to a physical model. Pressurization and depressurization rates and minute ventilation (MV) were measured under 24 conditions including pressure support of 10 or 20 cmH2O, positive end expiratory pressure (PEEP) of 5 or 10 cmH2O, ventilator circuit with “high”, “intermediate” or “low” resistance, and cushion deflated or inflated. In a clinical study pressurization and depressurization rates, MV and patient–ventilator interactions were compared in six patients with acute respiratory failure during conventional versus an “optimized” set-up (PEEP increased to 10 cmH2O, low resistance circuit and cushion inflated).

Results

In the bench study, all adjustments simultaneously applied (increased PEEP, inflated cushion and low resistance circuit) increased pressurization rate (46.7 ± 2.8 vs. 28.3 ± 0.6 %, p < 0.05), depressurization rate (82.9 ± 1.9 vs. 59.8 ± 1.1 %, p ≤ 0.05) and patient MV (8.5 ± 3.2 vs. 7.4 ± 2.8 l/min, p < 0.05), and decreased leaks (17.4 ± 6.0 vs. 33.6 ± 6.0 %, p < 0.05) compared to the basal set-up. In the clinical study, the optimized set-up increased pressurization rate (51.0 ± 3.5 vs. 30.8 ± 6.9 %, p < 0.002), depressurization rate (48.2 ± 3.3 vs. 34.2 ± 4.6 %, p < 0.0001) and total MV (27.7 ± 7.0 vs. 24.6 ± 6.9 l/min, p < 0.02), and decreased ineffective efforts (3.5 ± 5.4 vs. 20.3 ± 12.4 %, p < 0.0001) and inspiratory delay (243 ± 109 vs. 461 ± 181 ms, p < 0.005).

Conclusions

An optimized set-up for helmet NIV that limits device compliance and ventilator circuit resistance as much as possible is highly effective in improving pressure support delivery and patient–ventilator interaction.

Keywords

References

  1. Villa F, Cereda M, Colombo E, Berra L, Nacoti M, Foti G, Pesenti A (1999) Evaluation of four non invasive CPAP systems. Intensive Care Med S 66:A246
  2. Tonnelier JM, Prat G, Nowak E, Goetghebeur D, Renault A, Boles JM, L’her E (2003) Non-invasive continuous positive airway pressure ventilation using a new helmet interface: a case–control prospective pilot study. Intensive Care Med 29:2077–2080
  3. Antonelli M, Conti G, Pelosi P, Gregoretti C, Pennisi MA, Costa R, Severgnini P, Chiaranda M, Proietti R (2002) New treatment of acute hypoxemic respiratory failure: noninvasive pressure support ventilation delivered by helmet. A pilot controlled trial. Crit Care Med 30:602–608
    • View reference on PubMed
    • View reference on publisher's website
  4. Antonelli M, Pennisi MA, Pelosi P, Gregoretti C, Squadrone V, Rocco M, Cecchini L, Chiumello D, Severgnini P, Proietti R, Navalesi P, Conti G (2004) Noninvasive positive pressure ventilation using a helmet in patients with acute exacerbation of chronic obstructive pulmonary disease. A feasibility study. Anesthesiology 100:16–24
    • View reference on PubMed
    • View reference on publisher's website
  5. Rocco M, Dell’Utri D, Morelli A, Spadetta G, Conti G, Antonelli M, Pietropaoli P (2004) Noninvasive ventilation by helmet or face mask in immunocompromised patients. A case–control study. Chest 126:1508–1515
    • View reference on PubMed
    • View reference on publisher's website
  6. Conti G, Cavaliere F, Costa R, Craba A, Catarci S, Festa V, Proietti R, Antonelli M (2007) Noninvasive positive-pressure ventilation with different interfaces in patients with respiratory failure after abdominal surgery: a matched-control study. Respir Care 52:1463–1471
    • View reference on PubMed
  7. Navalesi P, Costa R, Ceriana P, Carlucci A, Prinianakis G, Antonelli M, Conti G, Nava S (2007) Non-invasive ventilation in chronic obstructive pulmonary disease patients: helmet versus facial mask. Intensive Care Med 33:74–81
  8. Crimi C, Noto A, Princi P, Esquinas A, Nava S (2010) A European survey of noninvasive ventilation practices. Eur Respir J 36:362–369
    • View reference on PubMed
    • View reference on publisher's website
  9. Principi T, Pantanetti S, Catani F, Elisei D, Gabbanelli V, Pelaia P, Leoni P (2004) Noninvasive continuous positive airway pressure delivered by helmet in hematological malignancy patients with hypoxemic acute respiratory failure. Intensive Care Med 30:147–150
  10. Chiumello D, Pelosi P, Carlesso E, Severgnini P, Aspesi M, Gamberoni C, Antonelli M, Conti G, Chiaranda M, Gattinoni L (2003) Noninvasive positive pressure ventilation delivered by helmet vs. standard face mask. Intensive Care Med 29:1671–1679
  11. Racca F, Appendini L, Gregoretti C, Stra E, Patessio A, Donner CF, Ranieri VM (2005) Effectiveness of mask and helmet interfaces to deliver noninvasive ventilation in a human model of resistive breathing. J Appl Physiol 99:1262–1271
    • View reference on PubMed
    • View reference on publisher's website
  12. Vargas F, Thille A, Lyazidi A, Campo FR, Brochard L (2009) Helmet with specific settings versus facemask for noninvasive ventilation. Crit Care Med 37:1921–1928
    • View reference on PubMed
    • View reference on publisher's website
  13. Moerer O, Fischer S, Hartelt M, Kuvaki B, Quintel M, Neumann P (2006) Influence of two different interfaces for noninvasive ventilation compared to invasive ventilation on the mechanical properties and performance of a respiratory system. A lung model study. Chest 129:1424–1431
    • View reference on PubMed
    • View reference on publisher's website
  14. Georgopoulos D, Prinianakis G, Kondili E (2006) Bedside waveforms interpretation as a tool to identify patient–ventilator asynchronies. Intensive Care Med 32:34–47
  15. Mojoli F, Venti A, Pozzi M, Via G, Braschi A (2009) Patient-ventilator interaction during pressure support ventilation: how to monitor and to improve it. Minerva Anestesiol 75:533–536
  16. Patroniti N, Foti G, Manfio A, Coppo A, Bellani G, Pesenti A (2003) Head helmet versus face mask for non-invasive continuous positive airway pressure: a physiological study. Intensive Care Med 29:1680–1687
  17. Taccone P, Hess D, Caironi P, Bigatello LM (2004) Continuous positive airway pressure delivered with a “helmet”: effects on carbon dioxide rebreathing. Crit Care Med 32:2090–2096
    • View reference on PubMed
    • View reference on publisher's website
  18. Mojoli F, Iotti GA, Gerletti M, Lucarini C, Braschi A (2008) Carbon dioxide rebreathing during non-invasive ventilation delivered by helmet: a bench study. Intensive Care Med 34:1454–1460
  19. Vignaux L, Tassaux D, Jolliet P (2007) Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study. Intensive Care Med 33:1444–1451
  20. Fabry B, Guttmann J, Eberhard L, Bauer T, Haberthiur C, Wolff G (1995) An analysis of desynchronization between the spontaneously breathing patient and ventilator during inspiratory pressure support. Chest 107:1387–1394
    • View reference on PubMed
    • View reference on publisher's website
  21. Schettino GP, Tucci MR, Sousa R, Barbas CS, Amato MB, Carvalho CR (2001) Mask mechanics and leak dynamics during noninvasive pressure support ventilation: a bench study. Intensive Care Med 27:1887–1891
  22. Cavaliere F, Conti G, Costa R, Proietti R, Sciuto A, Masieri S (2004) Noise exposure during noninvasive ventilation with a helmet, a nasal mask and facial mask. Intensive Care Med 30:1755–1760

Sign In

Connect with ICM

Top 5 Articles Editors Picks Supplement