In-line filter included into the syringe infusion pump assembly reduces flow irregularities

B. Brotschi| B. Grass| M. Weiss| C. Doell| V. Bernet
Physiological and Technical Notes
Volume 38, Issue 3 / March , 2012

Pages 518 - 522

Abstract

Purpose

To evaluate whether an in-line filter inserted in the syringe pump infusion line assembly influences start-up times and flow irregularities during vertical pump displacement at low infusion rates.

Methods

Fluid delivery after syringe pump start-up and after vertical displacement of the syringe pump by −50 cm was determined gravimetrically at flow rates of 0.5, 1.0 and 2.0 ml h−1. Measurements were repeated for each flow rate four times with two different syringe pumps with and without an in-line filter incorporated. Data are shown as median and range.

Results

Start-up times were reduced by an in-line filter at 0.5 ml h−1 flow rate from 355.5 s (0–660) to 115 s (0–320), whereas the effect was attenuated at higher flow rates. Pooling of fluid into the infusion system after lowering the infusion syringe pump was halved in all flow rates tested. Amount of infusion bolus after elevating the syringe pump by 50 cm was not affected by an in-line filter.

Conclusion

In the evaluated model in-line filters help to reduce flow irregularities and delay in drug delivery of syringe pumps at low flow rates and represent an option to optimize continuous administration of highly concentrated short-acting drugs at very small infusion rates.

Keywords

References

  1. Rooke GA, Bowdle TA (1994) Syringe pumps for infusion of vasoactive drugs: mechanical idiosyncrasies and recommended operating procedures. Anesth Analg 78:150–156
    • View reference on PubMed
    • View reference on publisher's website
  2. Lonnqvist PA, Lofqvist B (1997) Design flaw can convert commercially available continuous syringe pumps to intermittent bolus injectors. Intensive Care Med 23:998–1001
  3. Kern H, Kuring A, Redlich U, Dopfmer UR, Sims NM, Spies CD, Kox WJ (2001) Downward movement of syringe pumps reduces syringe output. Br J Anaesth 86:828–831
    • View reference on PubMed
    • View reference on publisher's website
  4. Neff T, Fischer J, Fehr S, Baenziger O, Weiss M (2001) Evaluation of the FASTSTART mode for reducing start-up delay in syringe pump infusion systems. Swiss Med Wkly 131:219–222
    • View reference on PubMed
  5. Neff T, Fischer J, Fehr S, Baenziger O, Weiss M (2001) Start-up delays of infusion syringe pumps. Paediatr Anaesth 11:561–565
    • View reference on PubMed
    • View reference on publisher's website
  6. Weiss M, Fischer J, Neff T, Baenziger O (2000) The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps. Anaesthesia 55:1094–1098
    • View reference on PubMed
    • View reference on publisher's website
  7. Neff TA, Fischer JE, Schulz G, Baenziger O, Weiss M (2001) Infusion pump performance with vertical displacement: effect of syringe pump and assembly type. Intensive Care Med 27:287–291
  8. Weiss M, Baenziger O, Neff T, Fanconi S (2000) Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med 26:776–779
  9. Weiss M, Hug MI, Neff T, Fischer J (2000) Syringe size and flow rate affect drug delivery from syringe pumps. Can J Anesth 47:1031–1035
    • View reference on PubMed
    • View reference on publisher's website
  10. Cook RI (2000) Syringe pump assemblies and the natural history of clinical technology. Can J Anesth 47:929–935
    • View reference on PubMed
    • View reference on publisher's website
  11. Leff RD, True WR, Roberts RJ (1987) A gravimetric technique for evaluating flow continuity from two infusion devices. Am J Hosp Pharm 44:1388–1391
    • View reference on PubMed
  12. Trautmann M, Zauser B, Wiedeck H, Buutenschön K, Marre R (2001) Bacterial colonization and endotoxin contamination of intravenous fluids. J Hosp Infect 37:225–236
    • View reference on publisher's website
  13. Jack T, Brent BE, Boehne M, Müller M, Sewald K, Braun A, Wessel A, Sasse M (2010) Analysis of particulate contaminations of infusion solutions in a pediatric intensive care unit. Intensive Care Med 36:707–711
  14. Kunac DL, Ball PA, Broadbet RS (1999) In-line intravenous filtration in neonates: help not hindrance. Aust J Hosp Pharm 29:321–327
  15. Ball PA (2003) Intravenous in-line filters: filtering the evidence. Curr Opin Clin Nutr Metab Care 6:319–325
    • View reference on PubMed

Sign In

Connect with ICM

Top 5 Articles Editors Picks Supplement