Accuracy of a transcutaneous carbon dioxide pressure monitoring device in emergency room patients with acute respiratory failure

P.-E. Gancel| E. Roupie| L. Guittet| S. Laplume| N. Terzi
Physiological and Technological Notes
Volume 37, Issue 2 / February , 2011

Pages 348 - 351

Abstract

Purpose

Transcutaneous CO2 monitors are widely used in neonatal ICUs. Until recently, these devices performed poorly in adults. Recent technical modifications have produced transcutaneous CO2 monitors that have performed well in adults with chronic illnesses. We evaluated the accuracy of one of these devices, the TOSCA® 500, in adults admitted to an emergency department for acute respiratory failure.

Methods

We prospectively collected 29 pairs of simultaneous transcutaneous arterial CO2 (PtcCO2) and arterial CO2 (PaCO2) values in 21 consecutive adults with acute respiratory failure (acute heart failure, n = 6; COPD exacerbation, n = 8; acute pneumonia, n = 6; and pulmonary embolism, n = 1). Agreement between PaCO2 and PtcCO2 was evaluated using the Bland-Altman method.

Results

Mean arterial oxygen saturation was 90%, arterial oxygen tension ranged from 32 to 215 mmHg, and PaCO2 ranged from 23 to 84 mmHg. The mean difference between PaCO2 and PtcCO2 was 0.1 mmHg, and the Bland-Altman limits of agreement (bias ± 1.96 SD) ranged from −6 to 6.2 mmHg. None of the patients experienced adverse effects from heating of the device clipped to the earlobe.

Conclusion

PtcCO2 showed good agreement with PaCO2 in adults with acute respiratory failure.

Keywords

References

  1. Crawford A (2004) An audit of the patient’s experience of arterial blood gas testing. Br J Nurs 13:529–532
    • View reference on PubMed
  2. Zaloga GP (1990) Evaluation of bedside testing options for the critical care unit. Chest 97:185S–190S
    • View reference on PubMed
  3. Thorson SH, Marini JJ, Pierson DJ, Hudson LD (1983) Variability of arterial blood gas values in stable patients in the icu. Chest 84:14–18
    • View reference on publisher's website
    • View reference on PubMed
  4. Eberhard P (2007) The design, use, and results of transcutaneous carbon dioxide analysis: current and future directions. Anesth Analg 105:S48–S52
    • View reference on publisher's website
    • View reference on PubMed
  5. Rosner V, Hannhart B, Chabot F, Polu JM (1999) Validity of transcutaneous oxygen/carbon dioxide pressure measurement in the monitoring of mechanical ventilation in stable chronic respiratory failure. Eur Respir J 13:1044–1047
    • View reference on publisher's website
    • View reference on PubMed
  6. Sanders MH, Kern NB, Costantino JP, Stiller RA, Strollo PJ Jr, Studnicki KA, Coates JA, Richards TJ (1994) Accuracy of end-tidal and transcutaneous pco2 monitoring during sleep. Chest 106:472–483
    • View reference on publisher's website
    • View reference on PubMed
  7. Cuvelier A, Grigoriu B, Molano LC, Muir JF (2005) Limitations of transcutaneous carbon dioxide measurements for assessing long-term mechanical ventilation. Chest 127:1744–1748
    • View reference on publisher's website
    • View reference on PubMed
  8. Maniscalco M, Zedda A, Faraone S, Carratu P, Sofia M (2008) Evaluation of a transcutaneous carbon dioxide monitor in severe obesity. Intensive Care Med 34:1340–1344
  9. McVicar J, Eager R (2009) Validation study of a transcutaneous carbon dioxide monitor in patients in the emergency department. Emerg Med J 26:344–346
    • View reference on publisher's website
    • View reference on PubMed
  10. Parker SM, Gibson GJ (2007) Evaluation of a transcutaneous carbon dioxide monitor (“Tosca”) in adult patients in routine respiratory practice. Respir Med 101:261–264
    • View reference on publisher's website
    • View reference on PubMed
  11. Bendjelid K, Schutz N, Stotz M, Gerard I, Suter PM, Romand JA (2005) Transcutaneous pco2 monitoring in critically ill adults: clinical evaluation of a new sensor. Crit Care Med 33:2203–2206
    • View reference on publisher's website
    • View reference on PubMed
  12. Kagawa S, Otani N, Kamide M, Gisiger PA, Eberhard P, Severinghaus JW (2004) Initial transcutaneous PCO2 overshoot with ear probe at 42 degrees c. J Clin Monit Comput 18:343–345
    • View reference on publisher's website
    • View reference on PubMed
  13. Roupie EE (1997) Continuous assessment of arterial blood gases. Crit Care 1:11–14
    • View reference on publisher's website
    • View reference on PubMed

Sign In

Connect with ICM

Top 5 Articles Editors Picks Supplement