Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation

Aissam Lyazidi| Arnaud W. Thille| Guillaume Carteaux| Fabrice Galia| Laurent Brochard| Jean-Christophe M. Richard
Original
Volume 36, Issue 12 / December , 2010

Pages 2074 - 2080

Abstract

Purpose

During volume-controlled ventilation, part of the volume delivered is compressed into the circuit. To correct for this phenomenon, modern ventilators use compensation algorithms. Humidity and temperature also influence the delivered volume.

Methods

In a bench study at a research laboratory in a university hospital, we compared nine ICU ventilators equipped with compensation algorithms, one with a proximal pneumotachograph and one without compensation. Each ventilator was evaluated under normal, obstructive, and restrictive conditions of respiratory mechanics. For each condition, three tidal volumes (VT) were set (300, 500, and 800 ml), with and without an inspiratory pause. The insufflated volume and the volume delivered at the Y-piece were measured independently, without a humidification device, under ambient temperature and pressure and dry gas conditions. We computed the actually delivered VT to the lung under body temperature and pressure and saturated water vapour conditions (BTPS).

Results

For target VT values of 300, 500, and 800 ml, actually delivered VT under BTPS conditions ranged from 261 to 396 ml (−13 to +32%), from 437 to 622 ml (−13 to +24%), and from 681 to 953 ml (−15 to +19%), respectively (p < 0.01). Respiratory system mechanics and application of an inspiratory pause significantly affected actually delivered VT. Assuming a set VT of 6 ml/kg of predicted body weight, a difference of 1–2 ml/kg with actually delivered VT would be commonly observed.

Conclusion

The difference between preset VT and actually delivered VT is clinically meaningful and differs across modern ICU ventilators.

Keywords

References

  1. (2000) Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. The Acute Respiratory Distress Syndrome Network. N Engl J Med 342:1301-1308
  2. Cote CJ, Petkau AJ, Ryan JF, Welch JP (1983) Wasted ventilation measured in vitro with eight anesthetic circuits with and without inline humidification. Anesthesiology 59:442–446
    • View reference on publisher's website
    • View reference on PubMed
  3. Forbat AF, Her C (1980) Correction for gas compression in mechanical ventilators. Anesth Analg 59:488–493
    • View reference on publisher's website
    • View reference on PubMed
  4. Lyazidi A, Richard JC-M, Thille AW, Besson O, Brochard L (2008) Are tidal volume delivered by the ventilator and received by the patient similar? Intensive Care Med 34:A0756
  5. Lellouche F, Taille S, Maggiore SM, Qader S, L’Her E, Deye N, Brochard L (2004) Influence of ambient and ventilator output temperatures on performance of heated-wire humidifiers. Am J Respir Crit Care Med 170:1073–1079
    • View reference on publisher's website
    • View reference on PubMed
  6. Ferreira JC, Chipman DW, Kacmarek RM (2008) Trigger performance of mid-level ICU mechanical ventilators during assisted ventilation: a bench study. Intensive Care Med 34:1669–1675
  7. Thille AW, Lyazidi A, Richard JC, Galia F, Brochard L (2009) A bench study of intensive care unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators. Intensive Care Med 35:1368–1376
  8. Richard JC, Carlucci A, Breton L, Langlais N, Jaber S, Maggiore S, Fougere S, Harf A, Brochard L (2002) Bench testing of pressure support ventilation with three different generations of ventilators. Intensive Care Med 28:1049–1057
  9. Brochard L, Roudot-Thoraval F, Roupie E, Delclaux C, Chastre J, Fernandez-Mondejar E, Clementi E, Mancebo J, Factor P, Matamis D, Ranieri M, Blanch L, Rodi G, Mentec H, Dreyfuss D, Ferrer M, Brun-Buisson C, Tobin M, Lemaire F (1998) Tidal volume reduction for prevention of ventilator-induced lung injury in acute respiratory distress syndrome. The multicenter trail group on tidal volume reduction in ARDS. Am J Respir Crit Care Med 158:1831–1838
    • View reference on PubMed
  10. Stewart TE, Meade MO, Cook DJ, Granton JT, Hodder RV, Lapinsky SE, Mazer CD, McLean RF, Rogovein TS, Schouten BD, Todd TR, Slutsky AS (1998) Evaluation of a ventilation strategy to prevent barotrauma in patients at high risk for acute respiratory distress syndrome. Pressure- and volume-limited Ventilation Strategy Group. N Engl J Med 338:355–361
    • View reference on publisher's website
    • View reference on PubMed
  11. Brower RG, Shanholtz CB, Fessler HE, Shade DM, White P Jr, Wiener CM, Teeter JG, Dodd-o JM, Almog Y, Piantadosi S (1999) Prospective, randomized, controlled clinical trial comparing traditional versus reduced tidal volume ventilation in acute respiratory distress syndrome patients. Crit Care Med 27:1492–1498
    • View reference on publisher's website
    • View reference on PubMed
  12. Cannon ML, Cornell J, Tripp-Hamel DS, Gentile MA, Hubble CL, Meliones JN, Cheifetz IM (2000) Tidal volumes for ventilated infants should be determined with a pneumotachometer placed at the endotracheal tube. Am J Respir Crit Care Med 162:2109–2112
    • View reference on PubMed
  13. Castle RA, Dunne CJ, Mok Q, Wade AM, Stocks J (2002) Accuracy of displayed values of tidal volume in the pediatric intensive care unit. Crit Care Med 30:2566–2574
    • View reference on publisher's website
    • View reference on PubMed
  14. Neve V, Leclerc F, Noizet O, Vernoux S, Leteurtre S, Forget P, Sadik A, Riou Y (2003) Influence of respiratory system impedance on volume and pressure delivered at the Y piece in ventilated infants. Pediatr Crit Care Med 4:418–425
    • View reference on publisher's website
    • View reference on PubMed
  15. Fujita Y, Imanaka H, Fujino Y, Takeuchi M, Tomita T, Mashimo T, Nishimura M (2006) Effect of humidifying devices on the measurement of tidal volume by mechanical ventilators. J Anesth 20:166–172
    • View reference on publisher's website
    • View reference on PubMed

Sign In

Connect with ICM

Top 5 Articles Editors Picks Supplement